712A - Pilot+ Omni - Sip & Puff Out Of Calibration
The Omni+ requires calibration of its Sip & Puff input.
To Calibrate the Sip & Puff input use the following description:
1 - Press and hold the Select button on the Omni+ Controller whilst switching the unit on.
2 - Once the screen has lit up press the Mode button to select ?Sip & Puff Calibration?.
3 - The screen will display the command Hard Puff. A hard puff should now be made by the user and repeated to ensure they can achieve the pressure consistently.
4 - As the user is exhaling, or puffing, press the Select button to store the value.
5 - The screen will then prompt the same procedure to calibrate the Soft Puff, Hard Sip and Soft Sip pressures.
6 - Once this sequence has been completed the Omni+ will prompt you to switch it off and on again.
The Calibration is now complete.